A 5-litre volume standard made of glass was calibrated. The method of determination and the uncertainty analysis is described. Three different methods for the analysis are compared; the ISO-standard 5168, the Swedish Standard SS-014150 and the recommendation from the BIPM. This calibration work comprises the contribution of the Swedish National Testing and Research Institute (SP) to an international comparison amongst several European national calibration laboratories. The determination was performed with water as test liquid and calculated for a reference temperature of 20 °C. The amount of water, related to the "dry" and "wet" volume respectively, was calculated from its measured mass and its corresponding density. The values of 4,99998 ± 3,5·10^-4 and 4,99763 ± 4,3·10^-4 litres are in good agreement with those from the other national laboratories.
SP REPORT 1992:33