A 5-litre volume standard was calibrated by eight European national calibration laboratories with the aim to detect the degree of agreement for this work. The main part of the calibration consisted of the determination of the dry and wet volume i.e. the volume that is contained and the volume that can be delivered from the standard. The report describes the intercomparison and compares the reported values as well as the estimated uncertainties. The eight results for both the dry (4,99993 l) and the wet volume (4,99764 l) lie within 0,02 percent, which is very satisfying. The intercomparison reproducibility can be stated as 0,021 and 0,028 percent for the two volumes, which is roughly twice the average laboratory repeatability (95 % confidence level). Although the judgement of the uncertainty contributions and their combinations differ in detail, the overall figures of the laboratories, compared on a common basis, correspond quite well.
SP REPORT 1991:29