Western European Metrology Club (WEMC) appointed a working group for Calibration Service to organize circulation of audit devices between laboratories in participating countries in order to increase mutual confidence in national calibration services. This report presents the results of the thermal audit circulation Th 3 conducted between September 1979 and April 1980. The audit pack consisted of a type S platinum-rhodium/platinum thermocouple whose thermoelectric voltage was measured at 600°C and 1000°C with reference junctions at 0°C. Participating laboratories in Sweden, Norway and Denmark compared calibration results and reported uncertainties. The overall agreement between laboratories was found to be very satisfactory, with measurements at 600°C agreeing within approximately ±0.5°C and measurements at 1000°C within approximately ±0.8°C.
SP-RAPP 1980:19