Flow Rate Ratio (FRR) Round Robin test
RISE., SP – Sveriges Tekniska Forskningsinstitut, SP – Sveriges Provnings- och Forskningsinstitut.
RISE., SP – Sveriges Tekniska Forskningsinstitut, SP – Sveriges Provnings- och Forskningsinstitut.
RISE., SP – Sveriges Tekniska Forskningsinstitut, SP – Sveriges Provnings- och Forskningsinstitut.
1995 (engelsk)Rapport (Fagfellevurdert)
Abstract [en]

Six companies/research institutes have participated in a Round Robin test concerning measurements of Flow Rate Ratio (FRR). A draft Nordtest method regarding FRR and polyethylene is partly based on the Round Robin. A flow rate ratio is obtained by dividing the flow rate at one condition by the flow rate at another condition. The reproducibility and the effect of melting time on the results has been investigated. Studies were also performed on the effects of the charged amount of material on the FRR results. All materials were polyethylenes with various densities and different amounts and types of additives. Two different loads at testing of each material has been used. The loads were selected together with the producer of the materials. For materials with high density two melting times were used, 4 minutes and 8 minutes, in order to find out if materials with high density sufficiently heated at the start of measurements. The Round Robin showed that a melting time of 4 minutes reached the same result as a melting time of 8 minutes. Consequently we see no reason to discriminate different densities of the polyethylenes. The amount of material that has been charged into the cylinder has been annotated during the test. The results show the amount of material does effect the result. However to draw any conclusions based on this Round Robin is not possible. No univocal trend is found. MFR respectively FRR have good reproducibility. For material with high MFR values the deviation tend to be bigger than for lower values. MFR obtained using a small load is often close to zero shear viscosity of the material. FRR is favorable for separating material that has just about the same MFR.

sted, utgiver, år, opplag, sider
Borås: Swedish National Testing and Research Institute (SP) , 1995. , s. 38
Serie
SP Rapport, ISSN 0284-5172 ; 1995_23
Emneord [en]
Flow Rate Ratio (FRR), Melt Flow Rate (MFR), polyethylene, Round Robin
HSV kategori
Identifikatorer
URN: urn:nbn:se:ri:diva-82833ISBN: 91-7848-552-5 (tryckt)OAI: oai:DiVA.org:ri-82833DiVA, id: diva2:2105172
Merknad

SP Rapport 1995:23

Tilgjengelig fra: 2026-09-25 Laget: 2026-09-25 Sist oppdatert: 2026-09-25bibliografisk kontrollert

Open Access i DiVA

Rapport(6864 kB)45 nedlastinger
Filinformasjon
Fil FULLTEXT01.pdfFilstørrelse 6864 kBChecksum SHA-512
268e153e47fdcc1363a746803d362cb28da0cbeb41c90df52f03e0e7cfd2c7e740f009d7839547ed6c834bf0cc571e5bceb080fc0f49d7349d7bc381b8e4fe6d
Type fulltextMimetype application/pdf

Av organisasjonen

Søk utenfor DiVA

GoogleGoogle Scholar
Totalt: 45 nedlastinger
Antall nedlastinger er summen av alle nedlastinger av alle fulltekster. Det kan for eksempel være tidligere versjoner som er ikke lenger tilgjengelige

isbn
urn-nbn

Altmetric

isbn
urn-nbn
Totalt: 9026 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf
v. 2.50.0