Reliability of an injection-moulded two-piece zirconia implant with PEKK abutment after long-term thermo-mechanical loading
University of Leuven, Belgium; University Hospitals Leuven, Belgium.
Humboldt-Universität zu Berlin, Germany.
Université de Lyon, France.
University of Leuven, Belgium.
Show others and affiliations
2020 (English)In: Journal of The Mechanical Behavior of Biomedical Materials, ISSN 1751-6161, E-ISSN 1878-0180, Vol. 110, article id 103967Article in journal (Refereed) Published
Abstract [en]

Zirconia implants are appreciated in some clinical indications in light of their aesthetic appearance and good biocompatibility. The aim of this work was to evaluate the performance of a newly developed two-piece zirconia/polyether ketone ketone (PEKK) implant-abutment combination after long-term cyclic loading in a hydrothermal environment, using a new protocol adapted from two available ISO standards. Sixteen implants (n = 8/group) were embedded according to ISO 14801 and divided into two groups: implants in the Observational Group (OG) were cyclically loaded for 60 days (98 N, 10 million loading cycles, 2 Hz) in 85 °C water in a chewing simulator, while non-loaded/non-aged implants (as-received) constituted the Control Group (CG). After 4.7 million loading cycles, one OG implant fractured in the chewing simulator. The surviving implants were compared to CG implants by X-ray diffraction (XRD) to investigate potential ageing as suggested by ISO 13356, but also μ-Raman spectroscopy, Focused-Ion-Beam - Scanning-Electron-Microscopy (FIB-SEM), and load-to-fracture. Ageing was shown to have limited influence on the evaluated zirconia implant, with increased monoclinic content after loading/ageing being to a shallow transformed zone of ~2 μm at the implant surface. However, OG implants showed a significantly decreased fracture load of 751 ± 231 N (CG: 995 ± 161 N; p =.046). These values enable clinical application, but the fact that one failure was recorded during cyclic fatigue along with the significant decrease in strength after cyclic loading/ageing suggest that there may be room for further optimization of especially the PEKK abutment. Furthermore, good agreement was observed between the fracture modes of the implant that failed during the cyclic fatigue experiment and the in vivo failure of one implant during pre-clinical trials, validating the interest of the in vitro protocol used in this work to check the reliability of zirconia implant.

Place, publisher, year, edition, pages
Elsevier Ltd , 2020. Vol. 110, article id 103967
Keywords [en]
Ageing, Fatigue, PEKK, Two-piece implants, Zirconia, Abutments (bridge), Biocompatibility, Cyclic loads, Fracture, Injection molding, Ion beams, ISO Standards, Ketones, Medical applications, Scanning electron microscopy, Silicon compounds, Thermal fatigue, Clinical application, Control groups, Focused ion beam-scanning electron microscopies, Hydrothermal environment, Implant abutments, Implant surface, Injection moulded, Thermo-mechanical loading
National Category
Natural Sciences
Identifiers
URN: urn:nbn:se:ri:diva-45610DOI: 10.1016/j.jmbbm.2020.103967Scopus ID: 2-s2.0-85088219240OAI: oai:DiVA.org:ri-45610DiVA, id: diva2:1458247
Note

Export Date: 10 August 2020; Article; Funding details: Fonds Wetenschappelijk Onderzoek, FWO, 12S8418N; Funding details: ORF21801; Funding text 1: The work was supported by a grant of the Oral Reconstruction Foundation (Reference: ORF21801 ) and by the Research Foundation - Flanders ( FWO Vlaanderen ) in light of F. Zhang's post-doctoral fellowship ( 12S8418N ).

Available from: 2020-08-14 Created: 2020-08-14 Last updated: 2020-12-01Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textScopus

Authority records

Adolfsson, Erik

Search in DiVA

By author/editor
Adolfsson, Erik
By organisation
Manufacturing Processes
In the same journal
Journal of The Mechanical Behavior of Biomedical Materials
Natural Sciences

Search outside of DiVA

GoogleGoogle Scholar

doi
urn-nbn

Altmetric score

doi
urn-nbn
Total: 36 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf