Procedure for automated low uncertainty assessment of empty cavity mode frequencies in Fabry-Pérot cavity based refractometryVisa övriga samt affilieringar
2024 (Engelska)Ingår i: Optics Express, E-ISSN 1094-4087, Vol. 32, nr 3, s. 3959-3973Artikel i tidskrift (Refereegranskat) Published
Abstract [en]
A procedure for automated low uncertainty assessment of empty cavity mode frequencies in Fabry-Pérot cavity based refractometry that does not require access to laser frequency measuring instrumentation is presented. It requires a previously well-characterized system regarding mirror phase shifts, Gouy phase, and mode number, and is based on the fact that the assessed refractivity should not change when mode jumps take place. It is demonstrated that the procedure is capable of assessing mode frequencies with an uncertainty of 30 MHz, which, when assessing pressure of nitrogen, corresponds to an uncertainty of 0.3 mPa.
Ort, förlag, år, upplaga, sidor
Optica Publishing Group (formerly OSA) , 2024. Vol. 32, nr 3, s. 3959-3973
Nyckelord [en]
Cavity resonators, Fabry-Perot interferometers, Fiber optic sensors, nitrogen, Cavity-mode frequencies, Fabry-Perot cavity, Gouy phase, Laser frequency, Mode frequencies, Mode number, Phase number, Refractometry, Uncertainty, Uncertainty assessment, adolescent, adult, article, female, frequency, human, laser, male, normal human, pressure, surgery, therapy, Uncertainty analysis
Nationell ämneskategori
Fysik
Identifikatorer
URN: urn:nbn:se:ri:diva-71939DOI: 10.1364/OE.513708Scopus ID: 2-s2.0-85183822866OAI: oai:DiVA.org:ri-71939DiVA, id: diva2:1841055
2024-02-272024-02-272024-05-27Bibliografiskt granskad